详细说明

示波器差分探头 TA043

收藏
产品型号 TA043
  • 产品说明
  • 产品参数

100 MHz的700 V差分示波器探头10:1和100:1

The TA043 is an active differential probe suitable for high common-mode voltage measurement applications up to ±700 V (DC + peak AC). It can be used with signal speeds of up to 100 MHz.

This differential probe extends the functionality of standard single-ended input oscilloscopes to allow a safe and accurate method of making high-voltage differential measurements. Applications include making safe measurements in power circuit applications and acquisition of low-speed balanced differential signals found in serial communications buses



TA043 differential probe specifications
BandwidthDC to 100 MHz (–3 dB) 
Attenuation ratio 10:1 / 100:1 
Accuracy±2%
Input impedance4 MΩ || 7 pF each side to ground 
Input voltage
Differential range (10:1)
Differential range (100:1) 
±70 V (DC + peak AC) or 70 V RMS
±700 V (DC + peak AC) or 500 V RMS
Common mode range±700 V (DC + peak AC) or 500 V RMS
(10:1 and 100:1) 
Absolute max. voltage
(differential or common-mode)
1000 V RMS CAT III (10:1 and 100:1) 
Output voltage 
Swing±7V into 50 kΩ load
Offset (typical) <±5 mV 
Noise (typical)0.9 mV RMS
Source impedance (typical)50 Ω
CMRR (typical) 85 dB @ 50 Hz; 55 dB @ 1 MHz
Ambient operating Temperature–10 to 40 ℃
Ambient storage temperature–30 to 70 ℃
Power requirements
Standard4xAA cells or 6 V DC / 200 mA mains adaptor
or regulated 9 V DC / 120 mA mains adaptor
Length of input leads 30 cm
Length of BNC lead 90 cm
Weight500 g 
Dimension (LxWxH) 202 mm x 83 mm x 38 mm 



×